Input Calibration
164 UDA2182 Universal Dual Analyzer Product Manual January 2009
Step Action Screen
4
Press
Enter
CATION PH
Sample Cal
pH Offset 0.00
Rst pH Offset
CATION PH
Sample Cal
pH Offset 0.00
Rst pH Offset
Use
to select
Sample Cal
5
DO NOT Remove the
electrode from the
process.
6
Once the reading is stable,
press
Enter
Change to Sample Value”
Use
keys to change the
displayed value to match the
value on the portable meter.
“Enter to save, Exit to cancel”
7
Follow the prompts at the top
and bottom of the screen.
“Cal Complete”
To recalibrate, press “Enter”.
“Enter = recal, Exit = exit”
8
If the calibration fails, an error
message will be displayed
across the bottom stripe of the
screen.
Make necessary adjustments
and re-calibrate.
Error Messages:
See Table 12-2.