Outputs Calibration
182 UDA2182 Universal Dual Analyzer Product Manual January 2009
Step Action Screen
8
Press
Enter
OUTPUT CAL
Output 1
Output 2
Output 3
OUTPUT CAL
Output 1
Output 2
Output 3
Use
to select
an Analog Output to be
calibrated
9
Press
Enter
OUTPUT 1
20mA Offset
4mA Offset
0
0
Reset 20mA Offs
Reset 4mA Offs
OUTPUT 1
20mA Offset
4mA Offset
0
0
Reset 20mA Offs
Reset 4mA Offs
Use
to select
” 20 mA Offset”
10
Press
Enter
OUTPUT 1
20mA Offset
4mA Offset
-147
3
Reset 20mA Offs
Reset 4mA Offs
OUTPUT 1
20mA Offset
4mA Offset
-147
3
Reset 20mA Offs
Reset 4mA Offs
The right most digit will be
“blinking”.
11
To correct the value on the
meter:
Use the keys to
increment or decrement the
value of the digit
Use the keys to move
the cursor to the next digit.
Repeat as required to
achieve a 20mA reading on
the test meter
When all digits have been
changed, press “Enter” to
store the 20mA value.
Press “Exit” to cancel. The
previous value is retained.