Honeywell UDA2182 Instrukcja Użytkownika Strona 233

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 258
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 232
Appendices
January 2009 UDA2182 Universal Dual Analyzer Product Manual 221
problem is that the cursor is positioned too far to the left or right of the flat portion of
the curve, move the cursor back to the flat portion of the curve.
15.
Perform another Air Calibration to correct any changes that occurred during the PBT.
16.
If you reached this point, you have both a working probe and analyzer that are
calibrated to one another correctly.
Przeglądanie stron 232
1 2 ... 228 229 230 231 232 233 234 235 236 237 238 ... 257 258

Komentarze do niniejszej Instrukcji

Brak uwag